PERFORMANCE
Range
0-10µS/cm, 0-100µS/cm
Accuracy
< 0.6% of the measuring range
Reagent Consumption (approximate)
Cation Resin (pre-rinsed) and Soda Lime; replace cartridges every two months
Sensor Classification
Cationic exchange and conductivity sensors
Degassing Efficiency
93.5%
Data Storage
Extract via USB 2.0
Power Supply
AC 110 to 240 V +10 / -15%; 48 to 63 Hz or AC/DC 20 to 30 V; 48 t0 63 Hz
OPERATING CONDITIONS
Sample Temperature Range
32 – 122° F (0 – 50 °C)
Ambient Temperature Range
32 – 140° F (0 – 60 °C)
Inlet Sample Flowrate
.031 - .044 GPM (7-10L/h), 6mm Swagelok
Inlet Sample Pressure
Maximum of 145 psig (10 bar)
Outlet Sample
Atmospheric
SIGNAL OUTPUTS
Analog Outputs
5 x 0/4 – 20 mA for measured signals
Alarms
General Alarm, Low Degassing Efficiency Alarm, Resin Exhausted Alarm
Optional
RS422/RS485
MECHANICAL
Construction
Transmitter – high strength ABS, IP67
Wet section – stainless steel or PVC
Dimensions
29 x 37 x 10" (613 x 939 x 254mm)
Mounting
Wall mount or Panel mount
Finish
Corrosion resistant